toggle menu

Temperature & Humidity sensor for reference measurements - discontinued

DISCONTINUED

Temperature & Humidity sensor with highest precision for reference measurements

IMPORTANT: Temperature & Humidity sensor for reference measurements is discontinued

Download PDF
General
Housing material PVDF black
Mechanical sensor protection sensor Standard polyethylene dust filter
Operating temperature -20...80°C
Storage temperature -10...60°C (non-condensing)
Air temperature
Principle NTC
Measuring range -20 ... 80 °C
Unit °C
Accuracy 0 ... +70 °C: ±0.15 K;
-20 ... +80 °C: ±0.25 K
Relative humidity
Principle Resistive-electrolytic
Measuring range 0 ... 100 %
Unit %
Measurement accuracy incl. reproducibility and hysteresis +15 ... +30 °C typical ±0.8% rH;
0 ... +50 °C: typical ±1.0% rH;
-20 ... +80 °C: typical ±2.5% rH
additional uncertainty of the reference system of 0.2 ... 0.8%
Number of calibration points Humidity: 13 points over full measurement range | Temperature: 2 points over full measurement range
Download PDF