toggle menu

Handheld measuring device XP201 for reference humidity - discontinued

High-precision reference measurement standard for industrial humidity calibrations

DISCONTINUED
DISCONTINUED
DISCONTINUED
DISCONTINUED
DISCONTINUED
DISCONTINUED

High-precision reference measurement standard for industrial humidity calibrations. Suitable as humidity reference in climate chambers or humidity generators. Mini USB interface with software, online data collection.

IMPORTANT: Hand-held measuring device XP201 is discontinued

Please check the alternative: All-in-One Handheld Device XA1000

  • Relative humidity, temperature and air pressure
  • resistive-electrolytic / Humidity
  • High-precision measurement of temperature and relative humidity (0.8 % accuracy), display of calculations and statistical functions, high-quality case, resistive-electrolytic humidity sensor, 2m cable, batteries and DAkks calibration certificate included in delivery
  • USB (Cable and SmartGraph3 software included)

Hand-held measuring device XP201 is discontinued

Please check the alternative: All-in-One Handheld Device XA1000

Advantages of the Hand-held measuring device XP201 humidity reference device

We from Lufft created with the XP Series true X-perts for a range of various measurements. The four device types are specialists on different measuring ranges. Our hand-held devices can be used for daily usage in service, in air conditioning and environment technology, in the food industry or for industrial temperature calibration.

XP201 - the most precise hand-held measuring device (0.8 %) for relative humidity:

With the XP201 we developed the most precise hand-held measuring device (0,5 %) for humidity, which Lufft ever offered. It can be used as a high-precision reference measuring instrument, e.g. in climate chambers and humidity generators, and is especially made for the industrial use. It distinguishes itself by its excellent stability, very good repeatable measurements, absolute hysteresis latitude as well as the non-existent drift of measurement values even at very high humidity levels. The XP201 has a mini USB port and comes with software for online data collection. It's delivered in a high-quality wooden case.

A few advantages in detail:

  • High quality case
  • Highest humidity precision (0.8%)
  • DAkks calibration certificate included in delivery
  • Resistive-electrolytic sensor, 2m cable and batteries included in delivery 
  • Easy to handle and robust sensor connection
  • With automatic recognition
  • Saves measuring campaignes
  • Allows all climate data to be calculated and archieved
  • Evaluation software SmartGraph3 included

Example of use of the Hand-held measuring device XP201 humidity reference device

We from Lufft created with the XP Series true X-perts for a range of various measurements. The four device types are specialists on different measuring ranges. Our hand-held devices can be used for daily usage in service, in air conditioning and environment technology, in the food industry or for industrial temperature calibration.

XP201 - The X-pert of humidity measurement

The XP201 is the perfect choice for humidity reference in climate chambers or humidity generators. With the high-precision measurement of up to 0,5% for humidity it is suitable as reference measurement standard for industrial humidity calibrations. For traceability to national standards a DAkkS calibration certificate is attached in every delivery.

Download PDF
Sensor power supply 5.5 V ± 10 % DC, max. 200 mA
Integrated air pressure sensor
Measuring range 800 ... 1100 mbar
Accuracy at 25°C, 1013.25mbar 0.5 mbar
Long-term stability typ. 1 mbar/year
Measurement resolution 0.024 mbar
Measuring principle Piezoresistive
Calculated measurement categories for external temperature/humidity sensors Mathematical: MIN/MAX/AVG/HOLD
Temperature °C/°F
Rel. humidity %r.h
Rel. humidity of ice %r.h
Water vapour density (absolute humidity) g/m3
Dew point temperature °C/°F
Frost point temperature °C/°F
Mixing ratio at saturation (100%) g/kg
Volume fraction of water vapour /mass fraction of water vapour %
Wet-bulb temperature °C/°F
Ice-bulb temperature °C/°F
Specifi c Enthalpy (mass of air) kJ/kg
Saturation vapour pressure above ice/water hPa
Vapour particle pressure hPa
Air density kg/m3
General
Dimensions 170 x 62 x 34 mm
Weight Approx. 205 g
Housing material sensor PVDF black
Mechanical sensor protection sensor Standard polyethylene dust filter
Storage conditions
Permissible ambient temperature -10 ... 60 °C
Operating rel. humidity <90 % RH non - condensing
Operating conditions
Operating rel. humidity <90 % RH non - condensing
Operating temperature -20 ... 80 °C
Admissible height above absolute altitude 4000 m
Power supply
Power supply 4 Alkaline LR6 AA 1.5 V/USB 5 V
Active power consumption Approx. 400 mW
Battery life passive Approx. 1 year
Battery life active min. 24 hours
Data storage
Integrated data storage Up to 200 gauges taking approx. 1 mill. values
Interface
USB Cable and SmartGraph3 software included in delivery
Display
Definition of measured values 2 decimal places
Display
Control Touch screen, capacitive
Technology TFT, resolution 240 x 320, 65 k colours, very good contrast due to Piezoresistive technology
Surface, toughened glass Degree of hardness: 7, scratch - resistant
Relative Humidity
Principle Resistive - electrolytic
Measuring range 0 ... 100 %
Unit %
Measurement accuracy incl. reproducibility and hysteresis +15 ... +30 °C: typ. ±0.8% rH
0 ... +50 °C: typ. ±1.0% rH
-20 ... +80 °C: typ. ±2.5% rH
additional uncertainty of the reference system of 0.2 ... 0.8%
Number of calibration points :Humidity: 13 points over full measurement range | Temperature: 2 points over full measurement range
Temperature
Principle NTC
Measuring range -20 ... 80 °C
Unit °C
Accuracy 0 ... +70 °C: ±0.15 K
-20 ... +80 °C: ±0.25 K
Download PDF