OPUS 20E for external Sensors - discontinued
The OPUS20E offers the highest flexibility and is excellent value for money. It allows the connection of up to 4 external temperature and relative humidity sensors, as well as 2 further analogue sensors. Intelligent BUS sensors can be integrated...
Air flow and differential pressure sensors are typically connected to the OPUS20E via analogue inputs as opposed to the maximum of 4 external temperature or humidity sensors that can be integrated via a digital BUS protocol. In connection with its LAN capabilities, the OPUS20E is able to realize universal measurement networks in real time. For standard applications the SmartGraph 3 comes into play, and in order to fulfill the 21 CFR 11 guidelines the well established and proven MCPS8 software is available.
IMPORTANT: OPUS 20E is discontinued
- Measurement of Input voltage 0-1V, Current measurement - 2-wires, Current measurement - 3-wires, Thermocouple K, Thermocouple J, Thermocouple S, PT100
- Input voltage 0-1V, Current measurement - 2-wires, Current measurement - 3-wires, Thermocouple K, Thermocouple J, Thermocouple S, PT100
- LAN datalogger with built-in sensors and the highest precision and with up to 10 external channels/sensors per Opus20E.
- USB (Cable and Monitoring Software SmartGraph3 included)
Advantages of the indoor/inroom datalogger OPUS 20E
- Highest flexibility with up to 10 external channels/sensors per OPUS20E.
- LAN capabilities to realize universal measurement networks in real time
- Monitoring and evaluation software SmartGraph3 included
- Power over Ethernet (PoE) available as option
- Power supply via batteries as well as USB-Connection possible
- Firmware online updatable
- If 21 CFR 11 guidelines have to be fulfilled, the well-established and proven MCPS8 software is available
Examples of the indoor/inroom datalogger OPUS 20E
The only LAN datalogger with 4 interfaces is often used in the following applications:
- Climate monitoring in buildings
- control of all climate-sensitive production processes
- in electronic data-processing centres
- control cabinets
- wind turbines
- storage rooms
- museums
- many other climate-sensitive applications…
Product variants - OPUS 20E for external Sensors
The OPUS 20E for external Sensors is available in 4 product variants:
- OPUS20E for external Sensors (Article no. 8120.30)
- OPUS20E for external Sensors, Neutral = without Lufft branding (Article no. 8120.30N)
- OPUS20E for external Sensors, PoE = Power over Ethernet energy supply (Article no. 8120.31)
- OPUS20E for external Sensors, PoE, Neutral = Power over Ethernet energy supply, without Lufft branding (Article no. 8120.31N)
General | |
Dimensions | 180 x 78 x 32 mm |
Measuring interval | 10/30 s, 1/10/12/15/30 min, 1/3/6/12/24 h |
Construction | Plastic housing |
Operating time with battery | > 1 Year |
Data storage | 16 MB, 3,200,000 measured values |
LC-Display | Height 90 x 64 mm |
Weight | Approx. 250 g |
Included in delivery | PC - Windows Software SmartGraph 3 for graphical and numerical representation of measured values / Instructions/ data cable/ battery/ WAGO connector / DIN rail bracket |
Interface | USB, LAN |
Storage interval | 1/10/12/15/30 min, 1/3/6/12/24 h |
Power supply | 4 x LR6 AA Mignon, USB, 24V DC |
Operating temperature | -20...50 °C |
Bus interface | RS485 |
Input voltage 0-1V | |
Measuring range | 0 ... 1 V |
Unit | V |
Accuracy | +/- 200 uV +/- 0.1 % of measured value |
Resolution | 500 uV |
Current measurement - 2-wires | |
Measuring range | 4 ... 20 mA |
Unit | mA |
Accuracy | ± 4 uA ± 0.1 % of measured value |
Resolution | 5 uA |
Current measurement - 3-wires | |
Measuring range | 0 ... 20 mA |
Unit | mA |
Accuracy | ± 4uA ± 0.1 % of measured value |
Resolution | 5 uA |
Thermocouple K | |
Measuring range | -200 ... 1200 °C |
Unit | °C |
Accuracy | ±1 °C +/- 0.5 % of measured value at -200 °C ... 0 °C ±1 °C +/- 0.2 % of measured value at 0...1200 °C |
Resolution | 0.2 °C |
Thermocouple J | |
Measuring range | -200 ... 1200 °C |
Unit | °C |
Accuracy | ±1 °C +/- 0.5 % of measured value at -200 °C ... 0 °C ±1 °C +/- 0.2 % of measured value at 0...1200 °C |
Resolution | 0.2 °C |
Thermocouple S | |
Measuring range | -50 ... 1700 °C |
Unit | °C |
Accuracy | ± 1 °C ± 0.5 % of measured value at -50 °C ... 0 °C ± 1 °C ± 0.2 % of measured value at 0 °C ... 1,700 °C |
Resolution | 0.2 °C |
PT100 | |
Measuring range | -200 ... 500 °C |
Unit | °C |
Accuracy | ± 0.2 °C ± 0.1 % of measured value |
Resolution | 0.02 °C |